通过单片机与微机电系统(MEMS)传感器阵列的交互来对创伤性脑损伤进行检测
Interfacing a Microelectromechanical System (MEMS) Sensor Array for Traumatic Brain Injury Detection with a Microcontroller
关键词:光机电;传感器;单片机;微机电系统
摘 要:Traumatic brain injuries (TBIs) result from exposure to high accelerations and are a serious threat to Soldiers in close contact with improvised explosive devices as well as sports players who are frequently involved in collisions. To improve TBI detection, the U.S. Army Research Laboratory (ARL) has developed a sensor small enough to be mounted in the ear. The sensor consists of an array of 3-axis microelectromechanical system (MEMS) acceleration threshold switches with different sensitivities that move to contacts under acceleration and complete a circuit. Previously, the outputs, which were voltage levels, required an analog-to-digital converter, but the implementation of the mechanism introduced a delay of 100 microsecs samples. This delay has caused the loss of data from switch closures that last less than 100 microsecs, so the sensor was redesigned with digital outputs, and a new program was developed. Clocked signals were used to simulate sensor data, and tests showed improved delays of 3 microsecs to 10 microsecs.